The Drinking Water Multiparameter Analyzer features a 7-inch color touchscreen, photoelectric detection technology, GPS wireless communication, and standard Modbus protocol.
pH:
The pH value is determined by collecting the potential difference between a reference electrode and a glass hydrogen ion-selective electrode, along with the thermistor value. The temperature is obtained from the thermistor, and the potential is converted to a pH value using the Nernst equation:
E = E₀ − 2.303 × t × log(C_H⁺)
Residual Chlorine:
A constant potential is applied between the measuring and reference electrodes. Two platinum electrodes (working and counter electrodes) form a micro-current measurement system with the reference electrode. Hypochlorous acid undergoes a redox reaction on the surface of the working electrode, and the resulting current is measured. The residual chlorine concentration is then calculated based on the correlation between current intensity and chlorine concentration. Simultaneously, temperature and pH of the water are measured and used for signal compensation to obtain an accurate residual chlorine concentration.
Low Turbidity:
A laser beam is directed vertically from air into the water, where it is scattered by suspended particles. A photodetector positioned at a 90-degree angle to the incident light captures the scattered light. The turbidity analyzer converts the intensity of the scattered light into a turbidity value based on a calibration curve that correlates light intensity with turbidity concentration.
| Modèle de produit | PWMA-2000 | |||
| Monitoring factors | Low turbidity | Chlore résiduel | pH | EC/TDS/Salinity (optional) |
| principle | 90-degree light scattering color method | Constant voltage method | Glass electrode method | Graphite electrode |
| Plage de mesure | (0-20)NTU | 0~20mg/L | 0~14 | 0~5000us/cm;0~5000mg/l |
| 0~100ppt | ||||
| répétabilité | ≤1% | ≤1% | ±0.1 | ≤1% |
| Zero point drift | ±0.015NTU | 0,05mg/L | ±0.1 | ≤1%F.S |
| résolution | 0.001NTU | 0,01mg/L | ±0.01 | 1.us/cm 2.mg/l 0.01ppt |
| Interactive | 7-inch color touch screen | |||
| Cleaning method | Automatic blowdown, electrochemical & pneumatic/hydraulic cleaning | |||
| Interface de communication | 4G communication, 1 RS485 (standard Modbus), 2 4-20mA, 1 switching input, 5 switching outputs | |||
| Data storage | Data local storage of 1 year of data, expandable TF card (optional), support data USB Excel table form derivative | |||
| Dimensions : | (230*350*580)mm | |||
| Matériau | Cabinet: carbon steel spraying; Monitoring Pool: Acrylic & Composite Plastic | |||
| Tension de fonctionnement | 220V AC 50Hz/24VDC | |||
| MTBF | ≥ 1440h/heure | |||
| Consommation d'énergie | 10W on average | |||
| Environnement de travail | Temperature: (0~50)°C; The inlet water pressure is less than 3kg, and the flow rate is 500~1600mL/min | |||
Prêt à améliorer vos opérations grâce à des solutions d'analyse en ligne de pointe ? Contactez-nous pour obtenir des conseils d'experts, des tarifs ou une démonstration personnalisée.