FPI CEMS provides application gas analysis solution, integrated with Particulate Monitoring Subsystem, TPF Monitoring Subsystem, Gaseous (Pollutant) Monitoring Subsystem, Data Acquisition and Processing Subsystem, covering all the indicators and parameters to meet different customers’ needs. Such systems are extensively installed in industrial boilers, indicators and other emission sources with real-time data uploaded to local authorities.
Mercury analyzer applies cold vapor atomic fluorescence technology (CVAFS), integrated with dilution sampling system to monitor the mercury presence in the stack in various forms. Automatic calibration of mercury check with internal CAL-200 calibration gas generator.
The system is extensively installed after desulfurization facility in coal-fired power plants and other fossil fuel boilers.The sample flue gas will be vaporized and analyzed in the optical cell without enrichment of particles, which gives a better response time. The CVAFS does not requires any adsorption and desorption on gold amalgam, which does not involve any loss of sample contents.
With decades of experience in CEMS research and development, FPI launched diverse and state-of-the-art analyzers which analyzers can be integrated into a complete continuous emission monitoring system. FPI can provide corresponding models for you to choose from below to meet different customers’ requirements for specific components monitoring.
Principe
CEMS-2000B Hg applies Dilution Extractive (40:1) & Cold Va-por Atomic Fluorescence Spectrometry (CVAFS) technologyThe mercury analyze emits an elementspecific spectrum, andthen measures the in-tensity of the emitted light to measurethe mercury concentration.
| Principes | Dilution extractive (40:1); Cold Vapor Atomic Fluorescence (CVAFS) |
|---|---|
| Composants | Hg⁰, Hg²⁺, Hgᵀ |
| Température | (-20~50)°C |
| Gamme | 0.1~20~500μg/m³ |
| Limite de détection | 0.05μg/m³ |
| Linéarité | ≤±2%F.S |
| Déviation de portée | ≤±2%F.S./24h |
| Dérive nulle | ≤±2%F.S./24h |
| Répétabilité | ≤2% |
| Temps de réponse | 180~360s |
| Alimentation électrique | 220VAC/50~60Hz |
| Sortie | 4~20mA, RS232 |
Prêt à améliorer vos opérations grâce à des solutions d'analyse en ligne de pointe ? Contactez-nous pour obtenir des conseils d'experts, des tarifs ou une démonstration personnalisée.
Tim Li
Responsable marketing
Tim Li
Bonjour, comment puis-je vous aider aujourd'hui ?