Capteur de conductivité

The conductivity online sensor is equipped with a built-in NTC temperature sensor and compensation algorithm, enabling automatic temperature compensation for measurement accuracy. The sensor utilizes an RS485 interface and supports the standard Modbus protocol.

Téléchargements et ressources

Obtenir un devis maintenant
Description

Présentation du produit

To measure the resistance or conductivity of a sample, the current and voltage across the electrodes of a conductivity cell must be measured. Conductance (G) is defined as the reciprocal of resistance:

G = 1 / R = I / V

The value of conductance depends on the geometry of the electrode plates and is represented by the cell constant (K), which is the ratio of the distance between the electrodes (L) to the surface area of the electrode plates (A):

K = L / A

Where:

  • K — Cell constant of the conductivity electrode
  • A — Effective area of the electrode plates
  • L — Distance between the electrode plates

This value is referred to as the cell constant, and it can be calculated from the electrode’s geometric dimensions. For example, when two square electrode plates each have an area of 1 cm² and are spaced 1 cm apart, the cell constant K = 1 cm⁻¹.
If the measured conductance (G) using this electrode is 1000 μS, then the conductivity of the solution is:

κ = 1000 μS/cm

Conductance can be converted into standardized conductivity (κ), which is independent of the electrode configuration. This is done by multiplying the conductance by the cell constant:

κ = G × L / A = G × K

Capteur de conductivité Spécifications techniques

Product model EC-200 EC-210
Operating Principle Two electrodes/four electrodes Two electrodes/four electrodes
Plage de mesure (0~5000/300000) us/cm Other measurement ranges can be customized; Temperature: (0~50)°C (0~5000/300000) us/cm Other measurement ranges can be customized; Temperature: (0~50)°C
répétabilité <1% <1%
Measurement accuracy ±1% ±1%
Drift (24h) <1%F.S <1%F.S
résolution 1us/cm 1us/cm
Temps de réponse 10s 10s
Niveau de protection IP68 IP68
Communication interface RS485, the standard Modbus protocol baud rate 9600 address bit default 50 RS485, the standard Modbus protocol baud rate 9600 address bit default 50
Dimensions (φ33*224)mm, cable 10 meters (can be customized) (φ16*146)mm, cable 10 meters (can be customized)
poids 0.5kg 0.1kg
Matériau 316L stainless steel/PVC/titanium alloy optional 316L stainless steel/titanium alloy
Mounting threads The upper end is NPT3/4 threaded, and the lower end can be equipped with NPT3/4 male thread M16*1 thread at the upper end (top side), M20*1.5 thread at the upper end (bottom side)
Operating voltage (9~36) V DC (9~36) V DC
Energy consumption 0.4 W 0.4 W
MTBF ≥1440h/time ≥1440h/time
Calibration period 1~3 months 1~3 months
Work environment (0~50)°C (0~50)°C

 

Applications polyvalentes dans tous les secteurs

  • Eaux de surface
  • Industrial wastewater
  • Eau de process industriel
  • Piscines

Principales caractéristiques de Capteur de conductivité

  • Stable cell constant, unaffected by polarization
  • Equipped with temperature compensation, suitable for a wide range of environments
  • Boîtier résistant à la corrosion et conforme à la norme d'étanchéité IP68 pour un fonctionnement sous-marin de longue durée
  • Interface de communication RS485 avec protocole Modbus standard pour une intégration facile

Pourquoi Capteur de conductivité Se démarque

Foire aux questions

Obtenir un devis maintenant

Débloquez l'analyse de précision avec Capteur de conductivité

Prêt à améliorer vos opérations grâce à des solutions d'analyse en ligne de pointe ? Contactez-nous pour obtenir des conseils d'experts, des tarifs ou une démonstration personnalisée.

Envoyez votre demande aujourd'hui

Remplissez le formulaire ci-dessous, et nous vous contacterons dans les plus brefs délais.

Coordonnées
Détails du projet