The system integrates X-ray fluorescence for metal analysis up to 30 elements and a beta-ray attenuation model for particulate mass monitoring. But module will be using the same sample moving sequentially across the detector and would be significantly effective in industry emission tracing.
The system integrates X-ray fluorescence for metal analysis up to 30 elements and a beta-ray attenuation model for particulate mass monitoring. But module will be using the same sample moving sequentially across the detector and would be significantly effective in industry emission tracing.
| Campo di misura | 0~100μg/m³ |
|---|---|
| Measuring Components (Optional) | Al, Si, P, S, Cl, K, Ca, Sc, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Ga, As, Se, Br, Rb, Sr, Zr, Mo, Pd, Ag, Cd, In, Sn, Sb, Te, Cs, Ba, Au, Hg, Tl, Pb, and 38 other kinds of elements. |
| Lower Detection Level (Except Al) | (0.01~10)ng/m³ (1 hour sampling) |
| Lower Detection Level of Al | <40ng/m³ |
| Other Elements Lower Detection Level | P<5ng/m³, S<4ng/m³, Cr<1ng/m³, Si<8ng/m³, Pb<1ng/m³, As<1ng/m³, Cd<3ng/m³ (1 hour sampling) |
| Precisione | <1% (Pb standard film) |
| Linearità | Pb, R>0.999 |
| Repeatability of Internal Standard Substances | <0.2% |
| Standard Film Accuracy | <±2% (Pb standard film) |
Particulate Matter
| Campo di misura | 0~1, 2, 5, 10mg/m³ |
|---|---|
| Limite di rilevamento | 5μg/m³ |
Analyzer Running Indicators
| Temperature Control Precision of X-ray Tube | <±0.1℃ |
| X-ray Tube Power | <15W |
| Running Cycle | 1 hour (default), optional |
| Sampling Flow Rate | 4~20L/min adjustable, recommend 16.7L/min |
| Errore di campionamento del flusso | <±2% |
| Tape Replacement Cycle | 2 months (in 1 hour running cycle) |
| X-ray Tube Replacement Cycle | >1 year |
| Cooling | Built-in temperature control |