AQMS-500 Schwefeldioxid-Analysator

FPI AQMS-500 sulfur dioxide analyzer applies UV fluorescence technology with photomultiplier tube (PMT) to measure SO2.

Downloads und Ressourcen

Jetzt ein Angebot einholen
Beschreibung

Produktübersicht

AQMS-500 measures the intensity of the characteristic fluorescence released by SO2 in an ambient air sample contained in the gas chamber when the air sample is irradiated by ultraviolet light passed through the chamber.

Removal of Interferences
The permeation scrubber acting as hydrocarbon kicker removes aromatic hydrocarbon such as xylene and na-phthalene which causes interference. Optical filtering are employed to improve the rejection of interference from high nitrogen oxides.
UV Source
The pulsing of the UV source lamp serves to increase the optical intensity whereby a greater UV energy throughput and lower detectable concentration arerealized.
PMT System
The characteristic fluorescence of SO2 is received by the photomultiplier tube and converted into an ele-ctrical signal. The number of electrons is increased by the photomultiplier system and the current or voltageis collected by the anode.
Data Storage and Analysis
Stored data are easily retrievable through the serial or Ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.

AQMS-500 Schwefeldioxid-Analysator Technische Daten

Grundsatz UV Fluorescence
Standard Bereich (0–500)ppb, expandable
Null Lärm ≤0.25ppb (RMS)
Spanne Rauschen ≤2.5ppb (RMS)
Anzeige Digital
Untere Nachweisgrenze 0,5ppb
Null Drift ±1ppb/24h
Span Drift <0,5% des Lesens
Linearität <1%F.S.
Präzision Precision(20%F.S): ≤2ppb
Precision(80%F.S): ≤4ppb
Reaktionszeit T90<120s
Durchflussmenge der Probe (650±65)sccm
Datenübertragung 2 channel analog (4~20)mA; 12 digital input/output; 4-way relay output;
Kalibrierung Mehrpunkt-Kalibrator
Ausgabe RS232/RS485/Ethernet
Betriebstemperatur 0~40°C
Luftfeuchtigkeit im Betrieb 0~95%RH(Keine Kondensation)
Leistungsbedarf (220±10%)V AC/(45~55)Hz, power≤300W
Abmessungen und Gewicht 178(H)x432(W)x609(D)nm, 22kg

Vielseitige Anwendungen in verschiedenen Branchen

Hauptmerkmale von AQMS-500 Schwefeldioxid-Analysator

  • Compliance with US EPA reference method
  • Continuous system diagnosis with alarm
  • Various outputs include RS232,RS485,Ethernet
  • Temperature and pressure compensation
  • User-friendly interface with large screen
  • Internal data logging with 1 min to 365 daysmultiple averages
  • Multi-tasking software allows viewingtest variables while operating
  • Critical orifices provide flow stability

Warum AQMS-500 Schwefeldioxid-Analysator Stellt sich heraus

Häufig gestellte Fragen

Jetzt ein Angebot einholen