AQMS-500 Sulfur Dioxide Analyzer

FPI AQMS-500 sulfur dioxide analyzer applies UV fluorescence technology with photomultiplier tube (PMT) to measure SO₂.

Downloads and Resources

Get A Quote Now
Description

Product Overview

AQMS-500 measures the intensity of the characteristic fluorescence released by SO₂ in an ambient air sample contained in the gas chamber when the air sample is irradiated by ultraviolet light passed through the chamber.

Removal of Interferences
The permeation scrubber acting as hydrocarbon kicker removes aromatic hydrocarbon such as xylene and na-phthalene which causes interference. Optical filtering are employed to improve the rejection of interference from high nitrogen oxides.
UV Source
The pulsing of the UV source lamp serves to increase the optical intensity whereby a greater UV energy throughput and lower detectable concentration arerealized.
PMT System
The characteristic fluorescence of SO₂ is received by the photomultiplier tube and converted into an ele-ctrical signal. The number of electrons is increased by the photomultiplier system and the current or voltageis collected by the anode.
Data Storage and Analysis
Stored data are easily retrievable through the serial or Ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.

AQMS-500 Sulfur Dioxide Analyzer Technical Specifications

Measurement Method UV Fluorescence
Measurement Range 0-500 ppb
Zero Noise ≤0.2 ppb
Span Noise ≤2.5 ppb
Lower Detection Limit ≤0.4 ppb
Linearity Error ±1 %F.S.
20 percent Range Precision ≤2 ppb
80 percent Range Precision ≤4 ppb
24h Zero Drift ±1 ppb
24h 20 percent Range Drift ±2.5 ppb
24h 80 percent Range Drift ±5 ppb
Response Time (Rise/Fall) ≤90 s
Voltage Stability ±0.4 %F.S.
Flow Stability ±10 %
Temperature Influence 0.2 ppb/℃
Long-term Zero Drift (7 days) ±2 ppb
Long-term Span Drift (7 days) ±5 ppb

Versatile Applications Across Industries

Key Features of AQMS-500 Sulfur Dioxide Analyzer

  • Compliance with US EPA reference method
  • Continuous system diagnosis with alarm
  • Various outputs include RS232,RS485,Ethernet
  • Temperature and pressure compensation
  • User-friendly interface with large screen
  • Internal data logging with 1 min to 365 daysmultiple averages
  • Multi-tasking software allows viewingtest variables while operating
  • Critical orifices provide flow stability

Why AQMS-500 Sulfur Dioxide Analyzer Stands Out

Frequently Asked Questions