AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H₂S to SO₂. Bypassing the sample through an SO₂ scrubber which removes any SO₂, the H₂S is converted to SO₂ using this internal H₂S converter and measured using the pulsed fluorescence technique. AQMS-550 H₂S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H₂S in ambient air at levels commonly required.
AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H₂S to SO₂. Bypassing the sample through an SO₂ scrubber which removes any SO₂, the H₂S is converted to SO₂ using this internal H₂S converter and measured using the pulsed fluorescence technique.
Grundsatz
AQMS-550 H₂S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H₂S in ambient air at levels commonly required.
| Grundsatz | UV-Fluoreszenz |
| Gemessenes Gas | H₂S in ambient |
| Bereich | (0-500) ppb, expandable |
| Maßeinheiten | ppb, ppm, μg/m³, mg/m³ (Selectable) |
| Untere Nachweisgrenze | ≤0.5 ppb (RMS) |
| Null Lärm | ≤0.25 ppb (RMS) |
| Spanne Rauschen | ≤2.5 ppb (RMS) |
| Null Drift | ±1 ppb/24hours |
| Span Drift | Span drift(20% F.S.): ±5 ppb Span drift(80% F.S.): ±10 ppb |
| Durchflussmenge der Probe | (650±65) sccm |
| Linearität | <1% F.S. |
| Reaktionszeit | T90≤120 s |
| Aufgangs-/Fallzeit | <30 second |
| Datenübertragung | 2 channel analog (4-20) mA; 2 analog (0-5) V; 12 digital input/output; 4-way relay output |
| Ausgabe | RS232, RS485, Ethernet |
| Betriebstemperatur | 0-40°C |
| Leistungsbedarf | (220±10%) V AC/(45-55) Hz, power≤350 W |
| Abmessungen | 178(H)×432(W)×609(D)mm |
| Gewicht | Host: ≤22 kg HTC conversion module: ≤10 kg |