AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H₂S to SO₂. Bypassing the sample through an SO₂ scrubber which removes any SO₂, the H₂S is converted to SO₂ using this internal H₂S converter and measured using the pulsed fluorescence technique. AQMS-550 H₂S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H₂S in ambient air at levels commonly required.
AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H₂S to SO₂. Bypassing the sample through an SO₂ scrubber which removes any SO₂, the H₂S is converted to SO₂ using this internal H₂S converter and measured using the pulsed fluorescence technique.
Princípio
AQMS-550 H₂S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H₂S in ambient air at levels commonly required.
| Princípio | Fluorescência UV |
| Gás medido | H₂S in ambient |
| Gama | (0-500) ppb, expandable |
| Unidades de medida | ppb, ppm, μg/m³, mg/m³ (Selectable) |
| Limite inferior detetável | ≤0.5 ppb (RMS) |
| Ruído zero | ≤0.25 ppb (RMS) |
| Ruído de alcance | ≤2.5 ppb (RMS) |
| Desvio zero | ±1 ppb/24hours |
| Desvio de alcance | Span drift(20% F.S.): ±5 ppb Span drift(80% F.S.): ±10 ppb |
| Caudal de amostragem | (650±65) sccm |
| Linearidade | <1% F.S. |
| Tempo de resposta | T90≤120 s |
| Tempo de subida/descida | <30 second |
| Transmissão de dados | 2 channel analog (4-20) mA; 2 analog (0-5) V; 12 digital input/output; 4-way relay output |
| Saída | RS232, RS485, Ethernet |
| Temperatura de funcionamento | 0-40°C |
| Necessidade de energia | (220±10%) V AC/(45-55) Hz, power≤350 W |
| Dimensões | 178(H)×432(W)×609(D)mm |
| Peso | Host: ≤22 kg HTC conversion module: ≤10 kg |