AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H₂S to SO₂. Bypassing the sample through an SO₂ scrubber which removes any SO₂, the H₂S is converted to SO₂ using this internal H₂S converter and measured using the pulsed fluorescence technique. AQMS-550 H₂S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H₂S in ambient air at levels commonly required.
AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H₂S to SO₂. Bypassing the sample through an SO₂ scrubber which removes any SO₂, the H₂S is converted to SO₂ using this internal H₂S converter and measured using the pulsed fluorescence technique.
原則
AQMS-550 H₂S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H₂S in ambient air at levels commonly required.
| 原則 | 紫外線蛍光 |
| 測定ガス | H₂S in ambient |
| 範囲 | (0-500) ppb, expandable |
| 測定単位 | ppb, ppm, μg/m³, mg/m³ (Selectable) |
| 検出下限値 | ≤0.5 ppb (RMS) |
| ゼロ・ノイズ | ≤0.25 ppb (RMS) |
| スパン・ノイズ | ≤2.5 ppb (RMS) |
| ゼロドリフト | ±1 ppb/24hours |
| スパン・ドリフト | Span drift(20% F.S.): ±5 ppb Span drift(80% F.S.): ±10 ppb |
| サンプル流量 | (650±65) sccm |
| 直線性 | <1% F.S. |
| 応答時間 | T90≤120 s |
| ライズ/フォールタイム | <30 second |
| データ伝送 | 2 channel analog (4-20) mA; 2 analog (0-5) V; 12 digital input/output; 4-way relay output |
| 出力 | RS232、RS485、イーサネット |
| 動作温度 | 0-40°C |
| 所要電力 | (220±10%) V AC/(45-55) Hz, power≤350 W |
| 寸法 | 178(H)×432(W)×609(D)mm |
| 重量 | Host: ≤22 kg HTC conversion module: ≤10 kg |