FPI AQMS-400 carbon monoxide (CO) analyzer measures ambient CO concentration by employing nondispersive infrared with gas filter correlation method technology.
FPI AQMS-400 carbon monoxide (CO) analyzer measures ambient CO concentration by employing nondispersive infrared with gas filter correlation method technology.
原則
Infrared energy emitted by light source is passed through gas chamber containing the air sample, and the quantitative absorption of energy by CO in the sample cell is measured by corresponding detector.
| Standard Range | (0~50)ppm, Expandable |
| 原則 | 非分散型赤外(NDIR) |
| ゼロ・ノイズ | ≤0.1ppm (RMS) |
| スパン・ノイズ | ≤0.25ppm (RMS) |
| 検出下限値 | ≤0.04ppm (RMS) |
| ディスプレイ | 0.05ppm |
| ゼロドリフト | ±0.1ppm |
| スパン・ドリフト | <0.5% of reading |
| 直線性 | <1%F.S. |
| 精密 | <1% |
| 応答時間 | T90<60s |
| サンプル流量 | (800±80)sccm |
| データ伝送 | 2 channel analog (4~20) mA; 12 digital input/output; 4-way relay output; |
| コミュニケーション | RS232/RS485/Ethernet |
| 動作温度 | US EPA Specification 20~30°C Actual applicable:-5~55°C |
| Operating Humidity | 0~95%RH |
| 校正 | Multi-point calibrator |
| Power Requirement | 220V AC/50Hz; 110V AC/60Hz |
| Dimensions and Weight | 178(H)x432(W)x597(D)mm, 28kg |
GFC
GFC (Gas filter correlation) technology is utilized to remove interference caused by moisture and other backgrounds.
Two gas filled chambers are mounted on a rotating disc, which pass through an IR beam alternately. The measure chamber is filled with nitrogen while the reference chamber is filled with high concentration CO. IR beam then passes through the sample gas cell. The difference in absorbance is measured and provides a direct output of the gas concentration.
Data storage and analysis
Stored data are easily retrievable through the serial or ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.