ホーム / 製品 / 残留塩素電極

残留塩素電極

The residual chlorine online sensor is an online constant-voltage type sensor. It features a built-in flow stabilization cell that effectively eliminates the impact of flow rate on sensor performance. Equipped with integrated temperature and pH sensors, it supports automatic compensation for water temperature and pH. The sensor adopts an RS485 interface and standard Modbus protocol for easy integration.

ダウンロードとリソース

今すぐ見積もりを取得
説明

製品概要

A constant potential is applied between the measuring and reference electrodes to form a micro-current measurement system consisting of two platinum electrodes (working and counter electrodes) and a reference electrode. Hypochlorous acid undergoes redox reactions on the surface of the working electrode, generating a current signal. The residual chlorine concentration is calculated based on the relationship between current intensity and chlorine concentration. Meanwhile, the system measures the water temperature and pH, and performs signal compensation to obtain an accurate residual chlorine concentration in the water.

残留塩素電極 技術仕様

製品モデル FCl-200
動作原理 Constant voltage glass electrode method
Measurement factors Residual chlorine pH temperature
測定範囲 (0-20) mg/L 0~14 (0~50)℃
測定精度 ±10%; When the concentration is less than 0.5mg/L± 0.05mg/L <0.1pH 0.1℃
pH compensation range 6~9
検出限界 0.02mg/L
決議 0.01mg/L
応答時間 ≤10s
通信インターフェース RS485, standard Modbus protocol
寸法: (250*203*165)mm
素材 Electrode: platinum ring glass; Detection cell: acrylic
動作電圧 24V DC
MTBF ≥ 1440h/回
消費電力 0.3w
Minimum flow rate 0.5-1.2L/min
洗浄方法: Compressed air/raw water recoil
校正期間 1 month
職場環境 (0~50)℃

多様な産業分野での応用

  • Finished water from waterworks and distribution network endpoints
  • Secondary water supply systems
  • Swimming pools

主な特徴 残留塩素電極

  • Fast response for real-time residual chlorine monitoring
  • Automatic pH and temperature compensation for wide application adaptability
  • Unique flow cell design ensures stable flow rate during measurement
  • 標準ModbusプロトコルのRS485通信インターフェイスで統合が容易
  • Replaceable electrodes, reagent-free operation, simple maintenance

なぜ 残留塩素電極 際立つ