AQMS-300 O₃ Analyzer(UV Photometric Absorption Technology)

FPI AQMS-300 Ozone analyzer measures ambient O3 concentration in ppb level by utilizing UV photometric absorption technology.

Downloads and Resources

Get A Quote Now
Description

Product Overview

The concentration of ambient ozone is proportional to UV light absorption since there is a significant characteristic absorption for ozone on wavelength of 254nm.
Periodical diversion on sample flow passing through the ozone scrubber will generate reference measurement, which is compared with sample measurement to provide stable and representative result.

AQMS-300 O₃ Analyzer(UV Photometric Absorption Technology) Technical Specifications

Measurement Method UV Photometry
Measurement Range 0-500 ppb
Zero Noise ≤ 0.1 ppb
Span Noise ≤ 1 ppb
Lower Detectable Limit ≤ 0.2 ppb
Linearity ± 1% F.S.
20% Range Precision ≤ 1 ppb
80% Range Precision ≤ 2 ppb
24h Zero Drift ± 0.5 ppb
24h 20% Range Drift ± 4 ppb
24h 80% Range Drift ± 4 ppb
Response Time (Rise/Fall) ≤ 20 s
Voltage Stability ± 0.4% F.S.
Flow Stability ≤ ± 10%
Ambient Temperature Interference < 0.2 ppb/℃
Long-term Zero Drift (7d) < ± 0.5 ppb
Long-term Span Drift (7d) < ± 4 ppb
Water Interference (2% H₂O) < ± 0.15% F.S.
NO₂Interference (500 ppb) < ± 0.15% F.S.

Versatile Applications Across Industries

Key Features of AQMS-300 O₃ Analyzer(UV Photometric Absorption Technology)

  • Accurate direct UV absorption with reference comparison;
  • Compliance with US EPA equivalent method;
  • Various outputs include ethernet and RS232;
  • User friendly interface with large screen;
  • Continuous system diagnosis with alarm;
  • Multi-tasking software allows viewing test variables while operating;
  • Temperature and pressure compensation;

Ozone scrubber
The ozone scrubber is filled with MnO2 as catalyst, which will convert O3 to O2 to form reference gas. Meanwhile, the presence of other components remains the same.

Data storage and analysis
Stored data are easily retrievable through the serial or ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.

Why AQMS-300 O₃ Analyzer(UV Photometric Absorption Technology) Stands Out

Frequently Asked Questions

Get A Quote Now

Unlock Precision Analysis with AQMS-300 O₃ Analyzer(UV Photometric Absorption Technology)

Ready to enhance your operations with industry-leading online analysis solutions? Get in touch for expert guidance, pricing, or a personalized demo.

Related Products

Send Your Inquiry Today

Fill out the form below, and we will be in touch shortly.

Contact Information
Project Details